ISO 20411:2018
p
ISO 20411:2018
67923

Status : Published (To be revised)

This standard will be replaced by ISO/AWI 20411
en
Format Language
std 1 96 PDF + ePub
std 2 96 Paper
  • CHF96
Convert Swiss francs (CHF) to your currency

Abstract

ISO 20411:2018 specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in pulse counting magnetic sector-type secondary ion mass spectrometers or quadrupole secondary ion mass spectrometers. It uses a test based on depth profile analysis of two isotopes in a reference material which has a gradual concentration change between low and high concentration regimes. It also includes a correction method for saturated intensity caused by the dead time of the detector. The correction can increase the intensity range for 95 % linearity so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction equations have been shown to be valid.

ISO 20411:2018 does not apply to time of flight mass spectrometers.

ISO 20411:2018 is only applicable to elements with minor isotopes. It is not applicable if the element is monoisotopic or contains isotopes with equal abundances.

Read sample 

Preview this standard in our Online Browsing Platform (OBP)

General information

  •  : Published
     : 2018-03
    : International Standard to be revised [90.92]
  •  : 1
     : 15
  • ISO/TC 201/SC 6
    71.040.40 