Abstract
This International Standard specifies a method for the secondary calibration of acoustic emission sensors as
receivers of elastic waves at the surface of a solid medium. The International Standard is applicable to laboratory
tertiary standard sensors and acoustic emission applications sensors.
The secondary calibration yields the frequency response of a sensor to waves of a type normally encountered in
acoustic emission work, namely Rayleigh waves. The source producing the signal used for the calibration is
mounted on the same surface of a test block as the sensor under test (SUT). The sensitivity of the sensor is
determined for excitation within the range of 100 kHz to 1 MHz. Sensitivity values are usually determined at
frequencies approximately 10 kHz apart. The units of the calibration are volts per unit of mechanical input
(displacement, velocity or acceleration).
Read sample
General information
-
Status: PublishedPublication date: 1999-07Stage: International Standard confirmed [90.93]
-