Тезис
ISO 16645:2016 is applicable to medical electron linear accelerators i.e. linear accelerators with nominal energies of the beam ranging from 4 MV to 30 MV, including particular installations such as robotic arm, helical intensity modulated radiotherapy devices and dedicated devices for intra operative radiotherapy (IORT) with electrons.
The cyclotrons and the synchrotrons used for hadrontherapy are not considered.
The radiation protection requirements and recommendations given in ISO 16645:2016 cover the aspects relating to regulations, shielding design goals and other design criteria, role of the manufacturers, of the radiation protection officer or qualified expert and interactions between stakeholders, radiations around a linear accelerator, shielding for conventional and special devices (including shielding materials and transmission values, calculations for various treatment room configurations, duct impact on radiation protection) and the radiological monitoring (measurements).
Общая информация
-
Текущий статус: ОпубликованоДата публикации: 2016-10
Исправленный вариант (английский): 2016-11Этап: Подтверждение действия международного стандарта [90.93] -
Версия: 1
-
Технический комитет :ISO/TC 85/SC 2ICS :13.280
- RSS обновления
Жизненный цикл
-
Сейчас
ОпубликованоISO 16645:2016
Стандарт, который пересматривается каждые 5 лет
Этап: 90.93 (Подтверждено)-
00
Предварительная стадия
-
10
Стадия, связанная с внесением предложения
-
20
Подготовительная стадия
-
30
Стадия, связанная с подготовкой проекта комитета
-
40
Стадия, связанная с рассмотрением проекта международного стандарта
-
50
Стадия, на которой осуществляется принятие стандарта
-
60
Стадия, на которой осуществляется публикация
-
90
Стадия пересмотра
-
95
Стадия, на которой осуществляется отмена стандарта
-
00